SOURce Subsystem
9
Keysight U2722A/U2723A Programmer’s Reference
53
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
Syntax
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <current>,
(@<ch>)
This command sets the current trigger level of the specified output channel. Units
are in amperes. The triggered level is a stored value that is transferred to the
output when an output step is triggered.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? (@<ch>)
Queries the stored value of triggered current.
Parameters
Returned Query Format
<NR3>
Examples
The following command sets the current trigger level to 0.02 A on channel 1.
CURR:TRIG 0.02, (@1)
The following query returns the current trigger level on channel 1.
CURR:TRIG? (@1)
Typical Response: +2.000000E-02
See Also
INITiate[:IMMediate]:TRANsient
[SOURce:]CURRent:RANGe
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Item
Type
Range of values
Default value
<ch>
NR1
1 through 3
Required parameter
<current>
NRf
The maximum value is dependent on the current range set
Required parameter
Содержание u2722a
Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Programmer s Reference ...
Страница 7: ...Keysight U2722A U2723A Programmer s Reference 7 List of Figures Figure 1 Status system diagram 21 ...
Страница 8: ...8 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 10: ...10 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 50: ...8 SENSe Subsystem 50 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 102: ...10 STATus Subsystem 102 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...14 Error Messages 136 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...