SENSe Subsystem
8
Keysight U2722A/U2723A Programmer’s Reference
45
SENSe:CURRent[:DC]:NPLCycles
Syntax
SENSe:CURRent[:DC]:NPLCycles <cycles>, (@<ch>)
This command sets the number of power line cycles taken for the current
measurement. The value is coupled to APERture by the equation:
APERture = NPLCycles/selected line frequency.
SENSe:CURRent[:DC]:NPLCycles? (@<ch>)
Query the number of power line cycles taken for the current measurement in the
defined channels.
Parameters
Returned Query Format
<NR1>
Examples
SENS:CURR:NPLC 10, (@1)
SENS:CURR:NPLC? (@1)
See Also
SENSe:CURRent[:DC]:APERture?
SENSe:VOLTage[:DC]:NPLCycles
Item
Type
Range of values
Default value
<ch>
NR1
1 through 3
Required parameter
<cycles>
NR1
0 to 255
Required parameter