SOURce Subsystem
9
Keysight U2722A/U2723A Programmer’s Reference
77
[SOURce:]VOLTage
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <voltage>,
(@<ch>)
Sets the actual voltage magnitude of the unswept output signal in terms of the
present operating units. The units are set to the default value, or alternately to a
different value under the UNIT subsystem. The AMPLitude may be used to specify
the level for either a time varying or non-time varying signal.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? (@<ch>)
Queries the sourced voltage.
Parameters
Returned Query Format
<NR3>
Examples
The following command sets the output voltage to 5 V on channel 1.
VOLT 5, (@1)
The following query returns the programmed voltage level on channel 1.
VOLT? (@1)
Typical Response: +5.00
See Also
[SOURce:]CURRent
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Item
Type
Range of values
Default value
<ch>
NR1
1 through 3
Required parameter
<voltage>
NRf
The maximum value is dependent on the voltage range set
Required parameter
Содержание u2722a
Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Programmer s Reference ...
Страница 7: ...Keysight U2722A U2723A Programmer s Reference 7 List of Figures Figure 1 Status system diagram 21 ...
Страница 8: ...8 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 10: ...10 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 50: ...8 SENSe Subsystem 50 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 102: ...10 STATus Subsystem 102 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 136: ...14 Error Messages 136 Keysight U2722A U2723A Programmer s Reference THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...