The LAN Reset button
This recessed button has two functions.
•
Pressing the button briefly invokes a preset of the instrument and
restores default measurement settings. This is equivalent to the
programming command
:SYSTem:PRESet
•
Pressing and holding the button for 3 seconds will reset the LAN
parameters to the factory default. This includes changing the
password for modifying the configuration via LAN back to the
default "
Keysight
". This is equivalent to the following sequence of
programming commands:
:SYSTem:COMMunicate:ETHernet:RESet
GPIB Interface
GPIB Interface
You can connect your GPIB interface into a star network, a linear
network, or a combination star and linear network. The limitations
imposed on this network are as follows:
•
The total cable length cannot exceed 20 meters.
•
The maximum cable length between devices is 4 meters, with an
average separation of 2m between devices over the whole bus.
•
No more than 15 devices may be interconnected on one bus.
Cables and Adapters
For further details on cables and adapters, refer to "Cables" section in
the User's Guide.
Getting Started
1
Keysight
N77xx Series
37