Installation Note N5242-90125 37
Description of the Upgrade
Installation Procedure for the Upgrade
Install the New Semirigid Test Set Cables
1.
Install the following cables in the order listed. Unless otherwise indicated,
use a 5/16-in torque wrench set to 10 in-lbs to tighten all cable
connectors.
For installation of the following cables, refer to
.
—
➀
–W167
(N5222-20120) A74 Bias T combiner, port 2 CPLR THRU
—
➁
–W179 (N5222-20124) A74 Bias T combiner to A32 port 2 test port
coupler port 2
For installation of the following cables, refer to
—
➀
–W161 (N5222-20119) A71 Bias T combiner, port 1 to CPLR THRU
—
➁
–W178 (N5222-20123) A71 Bias T combiner to A29 port 1 test port
coupler
Figure 19
Install A74 port 2 bias-T combiner semirigid cables (N5222-20120 and
N5222-20124)