20 Installation Note N5242-90125
Description of the Upgrade
Installation Procedure for the Upgrade
-
Refer to
for this step of the procedure. New parts are listed in
2.
Install the N5240-60097 (x1) cables as shown. Note the orientation of the
cable (item
➀
).
3.
Install the N5240-60098 (x1) cable as shown. Note the orientation of the
cable (item
➁
).
4.
Add 5023-3299 (x2) clamps and 0515-2141 (x2) clamp screws as shown
(item
➂
). Torque to 6 in-lbs.
5.
Mark bias combiner with fine permanent marker on MW-IN connector and
DC-RF-MW-OUT (item
➃
).
Figure 3
Install cables and clamps to bias tee combiners (N5240-60098,