Installation Note N5222-90129 47
Installation Procedure for the Upgrade
Step 33. Enable Options 417
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
— A keyboard and mouse must be connected to the network analyzer.
Option Enable Procedure
1.
Locate the email(s) from Keysight which contain license file attachments.
These emails are a result of
“Step 1. Obtain a Keyword and Verify the
.
2.
Copy the license file(s) from the email(s) to the
root directory
flash drive. More than one license file may be copied to the USB flash
3.
Insert the USB flash drive into the PNA’s USB drive slot.
4.
Verifying and editing the license file:
For these steps, refer to the example in
a.
Verify your USB flash drive is connected to a PC.
b.
Open your license file using a text read/write program similar to
Notepad.
c.
If you have more than one licensed feature, delete the feature that is
not
required for this upgrade. (e.g., in this case N5242B-423 is the
correct upgrade. So, N5242B-422 is to be deleted from the text file.)
For this step, you will need a USB flash drive.
A single license file may contain licenses for more than one instrument.