50 Installation Note N5222-90129
Installation Procedure for the Upgrade
-
on-line HELP. A list of equipment required to perform these adjustments is also
found in the service guide.
To view this service guide information, click the Chapter 3 bookmark “Tests
and Adjustments” in the PDF Service Guide
1
.
After the specified adjustments have been performed, the analyzer should
operate and phase lock over its entire frequency range.
EEPROM Backup
The analyzer uses arrays of correction constants to enable the analyzer to
produce accurate, leveled source signals and receive clean test signals. These
constants are stored in non-volatile EEPROM memory and in flash memory
files.
The adjustments listed here generate new correction constants. The analyzer
must have a backup of this new data in case any of the data becomes
corrupted.
To store the backup data, perform these steps:
— Navigate to the EEPROM Backup Utility, located at:
—
Windows 7 --
C:\Program Files (x86)\Keysight\Network
Analyzer\Service\eebackup.exe
— Windows 10 --
C:\Program Files\Keysight\Network
Analyzer\Service\eebackup.exe
— Run the program.
— Click Backup EEPROM.
— Click Backup TSMB Mem.
— Click Exit when the program has finished.
Operator’s Check
Perform the Operator’s Check to check the basic functionality of the analyzer.
For instructions, click the Chapter 3 bookmark “Tests and Adjustments” in the
PDF Service Guide
I
f you experience difficulty with the basic functioning of the analyzer, contact
“Contacting Keysight” on page 4
1. See