46 Installation Note N5222-90129
Installation Procedure for the Upgrade
-
Step 31. Reinstall the Outer Cover
For instructions, click the Chapter 7 bookmark “Removing the Covers” in the
PDF Service Guide
1
.
Step 32. Remove Option 217 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must be running.
— A keyboard and mouse must be connected to the network analyzer.
Option 217 License Removal Procedure
1.
To start the Keysight License Manager, press Start > Keysight License
Manager > Keysight License Manager . A Keysight License Manager dialog
box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Select Desired Option list, click
217
.
4.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
5.
A message displays stating that the option removal was successful.