Specifications and Characteristics
1
Keysight N1913/1914A Service Guide
39
For Keysight U2000 Series power sensors
In FAST mode (using Free Run trigger), for a 10 dB decreasing power step, the
settling time is
25 ms
[1]
.
[1] When a power step crosses the auto-range switch point of the sensor, add 25 ms.
Table 1-9
U2000 Series settling time
Number of
averages
1
2
4
8
16
32
64
128
256
512 1024
Settling time
[a]
(s)
(Normal mode)
[a] Manual filter, 10 dB decreasing power step (not across the switching point)
0.045 0.09 0.17 0.34 0.66
1.3
2.6
5.2
10.4 20.9 41.9
Settling time
(s)
(x2 mode)
0.042 0.05 0.09 0.17 0.34 0.66
1.3
2.6
5.2
10.4 20.9
Содержание N1913
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters Service Guide...
Страница 16: ...16 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...18 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 20: ...20 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 62: ...2 Performance Tests 62 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 66: ...3 Adjustments 66 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 78: ...5 Troubleshooting Guide 78 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 132: ...6 Repair Guide 132 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 140: ...7 Contacting Keysight Technologies 140 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...