Performance Tests
2
Keysight N1913/1914A Service Guide
49
Complete Equipment List
Instrument
Critical specifications
Recommended Keysight
model number
Alternative Keysight
model number
Meters
Power Meter
Dual channel
Absolute accuracy: 0.5%
Power reference accuracy: 0.9% – (a best
capability measurement is required for the
power reference output — the power level
must be accurately measured, and the
uncertainty of this measurement must also be
known)
E4419B
E4419A
Power Sensor
Frequency: 50 MHz
Amplitude range: –70 dBm to –20 dBm
SWR: 1.15 at 50 MHz
8481D
Power Sensor
Frequency: 50 MHz
Amplitude range: –30 dBm to +20 dBm
SWR: 1.1 at 50 MHz
N8482A
8482A
Power Sensor
Frequency: 50 MHz
Amplitude range: –70 dBm to +20 dBm
SWR: 1.06 maximum
E4412A
Attenuator
20 dB Fixed Attenuator
N-type (m,f)
8491B (Option 020)
Содержание N1913
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters Service Guide...
Страница 16: ...16 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...18 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 20: ...20 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 62: ...2 Performance Tests 62 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 66: ...3 Adjustments 66 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 78: ...5 Troubleshooting Guide 78 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 132: ...6 Repair Guide 132 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 140: ...7 Contacting Keysight Technologies 140 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...