6
Repair Guide
120
Keysight N1913/1914A Service Guide
Instructions
Visual
Step 1
Carefully lift and remove the calibrator plug. [This step is not
applicable to units with rear-panel connectors.]
This step is only applicable for calibrator assembly with the
calibrator plug.
For more information on the improved calibrator assembly,
“Improved calibrator assembly”
Step 2
Use Circlip pliers to remove N1912-61806 sensor flex
assembly.
Release the tab holding the flex cable to the display interface
board, and then disconnect it.
Step 3
Disconnect the white plastic plug from the display interface
board.
Calibrator plug
Содержание N1913