1
Specifications and Characteristics
40
Keysight N1913/1914A Service Guide
Figure 1-5
U2000 Series settling time with auto-filter
-50 dBm
+10 dBm
+2 dBm
-40 dBm
45 ms
90 ms
2.6 s
2.7 s
460 ms
2.8 s
39 s
42 s
42 s
-
4 dBm
-
10 dBm
-20 dBm
-30 dBm
Normal
mode
45 ms
82 ms
1.3 s
1.5 s
450 ms
1.6 s
20 s
24 s
24 s
x2
mode
Maximum sensor power
within a range
U2000/1/2/4A
U2000/1/2H
U2000/1B
+20 dBm
+40 dBm
+12 dBm
+32 dBm
+6 dBm
+26 dBm
0 dBm
-
10 dBm
-
20 dBm
-
30 dBm
-
40 dBm
+20 dBm
+10 dBm
0 dBm
-
10 dBm
-
20 dBm
Minimum sensor power
within a range
Sensor
dynamic
range
Typical
settling
times
0 dBm
0 dBm
+12 dBm
+10 dBm
Содержание N1913
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters Service Guide...
Страница 16: ...16 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...18 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 20: ...20 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 62: ...2 Performance Tests 62 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 66: ...3 Adjustments 66 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 78: ...5 Troubleshooting Guide 78 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 132: ...6 Repair Guide 132 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 140: ...7 Contacting Keysight Technologies 140 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...