22
N1000A Quick Start Guide
2
Installing the N1000A
Step 3. Position the N1000A
Position the N1000A so that it will have sufficient clearance for airflow around the
top, back, and sides. Review
Table 2
to confirm that your operating or storage
environment is suitable for the instrument.
Table 2
Environmental Specifications
Item
Specification
Use
For indoor use only
Temperature
Operating
10 °C to +40 °C (50 °F to + 104 °F)
Non-operating
–40 °C to +70 °C (–40 °F to +158 °F)
Maximum relative humidity
Type tested at 95% RH, +40 degrees C
(non-condensing)
Altitude (Operating)
Up to 4,600 meters (15,000 ft)
Line Power
*
* The instruments can operate with mains supply voltage fluctuations up to ±10% of the nominal voltage.
100/120Vac, 50/60/400 Hz
220/240Vac, 50/60 Hz
700 Watts Max.
Weight
Mainframe without Modules (
characteristic
)
19.5 kg (43 lb)
Modules (characteristic)
1.2 kg (2.6 lb)
Dimensions (excluding handle)
Without front connectors and rear feet
234 mm H x 426 mm W x 530 mm D
(9.23 inch x 16.76 inch x 20.9 inch)
With front connectors and rear feet
234 mm H x 426 mm W x 601 mm D
(9.23 inch x 16.76 inch x 23.67 inch)
With front cover and rear feet
234 mm H x 426 mm W x 612 mm D
(9.23 inch x 16.76 inch x 24.1 inch)
NOTE
The official and complete specifications are located in the N1000A’s help system.
Содержание N1000A DCA-X Series
Страница 1: ...Keysight N1000A DCA X Series Wide Bandwidth Oscilloscope Quick Start Guide...
Страница 7: ...Introduction 1 N1000A Quick Start Guide 7 Front Panel Features Figure 3 Front Panel Features...
Страница 8: ...8 N1000A Quick Start Guide 1 Introduction Rear Panel Features Figure 4 Rear Panel Features...
Страница 16: ...16 N1000A Quick Start Guide 1 Introduction...
Страница 36: ...36 N1000A Quick Start Guide 2 Installing the N1000A...
Страница 44: ...44 N1000A Quick Start Guide 5 Contacting Keysight...
Страница 46: ...46 N1000A Quick Start Guide W warm up time 33 weight 22 Windows XP second desktop 32 setup wizard 28...