26
N1000A Quick Start Guide
2
Installing the N1000A
Step 6. Connect the Power
• Connect the line cord as shown in the figure below. The N1000A automatically
adjusts for line input voltages. There is no voltage selection switch. The line
cord provided is matched by Keysight to the country in which the order
originates.
Figure 10
LINE Power Cord Connection to N1000A
CAUTION
Always use the three-prong AC power cord supplied with this instrument. Failure to
ensure adequate earth grounding by not using this cord may cause product damage.
CAUTION
Do not
connect ac power until you have verified the line voltage is correct as
described in
Table 2
on page 22. Damage to the equipment could result.
CAUTION
This instrument has auto-ranging line voltage input; be sure the supply voltage is
within the specified range and that voltage fluctuations do not to exceed 10 percent of
the nominal supply voltage.
CAUTION
Windows registry: If the N1000A is mounted in a rack or cabinet, do not use the
system power switch to disconnect power from the instrument. Instead, use the
N1000A front-panel power switch. Using the system power switch may corrupt the
Windows registry requiring you to perform the recovery procedure to restore normal
operation to the N1000A.
Содержание N1000A DCA-X Series
Страница 1: ...Keysight N1000A DCA X Series Wide Bandwidth Oscilloscope Quick Start Guide...
Страница 7: ...Introduction 1 N1000A Quick Start Guide 7 Front Panel Features Figure 3 Front Panel Features...
Страница 8: ...8 N1000A Quick Start Guide 1 Introduction Rear Panel Features Figure 4 Rear Panel Features...
Страница 16: ...16 N1000A Quick Start Guide 1 Introduction...
Страница 36: ...36 N1000A Quick Start Guide 2 Installing the N1000A...
Страница 44: ...44 N1000A Quick Start Guide 5 Contacting Keysight...
Страница 46: ...46 N1000A Quick Start Guide W warm up time 33 weight 22 Windows XP second desktop 32 setup wizard 28...