Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure
49
Calibration of Frequency Gain
Configuration: Frequency Gain
1
Configure each function and range shown in the table below.
2
Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see
3
Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the above
table.
NOTE
Before step 25 can be performed, the 2 VAC gain, flatness and zero (steps 13, 18,
and 22) must be performed. Review
on page 44 before beginning this procedure.
SFP Cal step
Step name
Calibrator input
freq (Hz)
Calibrator input
amplitude, function
25
Frequency Gain
10,300
2.00 VAC