Test 8: Capacitance Performance Verification Tests
32
Keysight M9182A/M9183A Calibration and Verification Procedure
Test 8: Capacitance Performance Verification Tests
1
Make sure you have read
2
Set the Capacitance function on the M9182A/83A DMM.
3
Connect the 1
μ
F capacitor to the input terminals of the DMM, as shown
below.
4
Select each range in the order shown below. Compare capacitance
measurement results from the M9182A/83A DMM to the appropriate test limits
shown in the table. Be certain to allow for appropriate source settling.
Input
capacitance
(F)
DMM
function
DMM
range (F)
Quick
check
Error from nominal (± F)
24 hours
90 days
1 year
1.0000E-06 Capacitance 1.000000E-06
11.000E-09 11.000E-09 11.000E-09
1.0000E-06 Capacitance 10.00000E-06
Q
20.00E-09
20.00E-09
20.00E-09