Test 1: DC Volts Verification Tests
Keysight M9182A/M9183A Calibration and Verification Procedure
19
A “Q” in the Quick Check column indicates a quick performance verification
test point. All measurements are taken at 32 NPLC
Calibrator
input voltage
(V)
DMM
function
DMM
range (V)
Quick
check
Error from nominal (± V)
24 hours
90 days
1 year
200.0000E-03
DCV
200.0000E-03
Q
7.0E-06
9.6E-06
12E-06
000.0000E+00
DCV
200.0000E-03
Q
1.0E-06
1.6E-06
2E-06
-200.0000E-0
3
DCV
200.0000E-03
7.0E-06
9.6E-06
12E-06
2.00
DCV
2.00
44E-06
64E-06
86E-06
000.0000E+00
DCV
2.00
4E-06
4E-06
6E-06
-2.0
0
DCV
2.00
Q
44E-06
64E-06
86E-06
20.00
DCV
20.00
920E-06
1.14E-03
1.56E-03
000.0000E+00
DCV
20.00
120E-06
140E-06
160E-06
-20.0
0
DCV
20.00
920E-06
1.14E-03
1.56E-03
200.0000E+00
DCV
200.0000E+00
Q
6.4E-03
8.2E-03
10.6E-03
000.0000E+00
DCV
200.0000E+00
400E-06
400E-06
600E-06
-200.0000E+0
0
DCV
200.0000E+00
6.4E-03
8.2E-03
10.6E-03
300.0000E+00
DCV
300.0000E+00
39.6E-03
69.9E-03
75.9E-03
000.0000E+00
DCV
300.0000E+00
600E-06
900E-06
900E-06
-300.0000E+0
0
DCV
300.0000E+00
39.6E-03
69.9E-03
75.9E-03