Test 10: Source IDC Performance Verification Tests
Keysight M9182A/M9183A Calibration and Verification Procedure
35
Source
calibrator (
Ω
)
DMM
function
DMM output
current (A)
Quick check
Error from nominal (± A)
24 hours
90 days
1 year
(N/A)
SourceIMeasV
10.00000E-03
182.50E-06
182.50E-06
182.50E-06
1000
SourceIMeasV
1.000000E-03
18.250E-06
18.250E-06
18.250E-06
10,000
SourceIMeasV
100.0000E-06
1.8250E-06
1.8250E-06
1.8250E-06
100,000
SourceIMeasV
10.00000E-06
245.00E-09
245.00E-09
245.00E-09
1,000,000
SourceIMeasV
1.000000E-06
24.500E-09
24.500E-09
24.500E-09