![Keysight M9181A Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/keysight/m9181a/m9181a_calibration-and-verification-procedure_1961485029.webp)
Test 6: AC Volts Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure
29
Test 6: AC Volts Verification Test
1
Make sure you have read
2
Connect the calibrator to the input terminals of the DMM, as shown below.
3
On the M9182A/83A DMM, set the AC Volts function and the frequency filter
to Slow (10 Hz). Each measurement takes 2.5 seconds to complete.
4
Select each range in the order shown in the following table. Use the calibrator
to provide the indicated input voltage and frequency. Note the last three steps
set the frequency filter to Fast (400 Hz).
5
Compare AC voltage measurement results from the M9182A/83A DMM to the
appropriate test limits shown in the table. Be certain to allow for appropriate
source settling.