Test 2: DC Current Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure
23
The following examples illustrate how to calculate the proper values in step 8,
above.
– If the value of the resistor is 10.02000e6 then the value programmed for the 20
μ
A range is 20e-6 × 10.02000e6 = 200.4 V.
– For the 2
μ
A range the value programmed is 2e-6 × 10.02000e6 = 20.04 V, and
for the 200 nA range the value programmed is 200e-9 × 10.02000e6 = 2.004 V.
WARNING
Any voltages greater than 42 V may present a safety hazard. Make certain
that these voltages are not accessible.
Calibrator
source voltage
Calibrator source
IRange (A)
DMM
function
DMM
range (A)
Quick
check
Error from nominal (± A)
24 hours
90 days
1 year
-200.0000E-06
DCI
200.0000E-06
Q
304.0E-09
440.0E-09
600.0E-09
Irange *
Measured 10M
20.00000E-06
DCI
20.00000E-06
10.40E-09
16.60E-09
26.80E-09
Irange *
Measured 10M
-2.000000E-06
DCI
2.000000E-06
1.080E-09
1.660E-09
4.360E-09
Irange *
Measured 10M
200.0000E-09
DCI
200.0000E-09
300.0E-12
366.0E-12
400.0E-12