Troubleshooting
http://www.keysight.com/find/M8920A
If the M8920A Radio Test Set is not communicating via the LAN connection,
check for the presence of blinking yellow LEDs on the rear panel LAN
connector. If the ACT LED is not blinking, check the LAN cable and LAN integrity.
The LAN connection is on the controller (either front panel of
the embedded controller, or wherever the LAN connection is
on the user-supplied controller.
To meet specifications, the module must be aligned. Refer to
Operation of the M8920A (page 67)
.
Perform an Alignment. Select System
> Alignments > Align Now > Align All
Now.
If the previously performed alignments did not resolve the problem, select the
System
, Restore Align Defaults All. Then select the System
, Align-
ments, Align Now, Align Now All.
Is the N9093 Radio Test mode displaying an error message? If so, refer to the
Check if the external frequency reference is selected but not available. Verify
that it is selected by choosing Input/Output, Freq Ref Input. The frequency of
the reference should be set correctly.
If you are using a Windows program other than the instrument application, you
may notice it running slow. Place the instrument application in Sweep, Sweep-
/Measurement, Single.
You can get automatic electronic notification of new firmware
releases and other product updates and information by
subscribing to the Keysight Technologies Test & Measurement E-
Mail Notification Service for your Radio Test Solution at
http://keysight.com/find/MyKeysight
80
M8920A Getting Started Guide
Содержание M8920A
Страница 1: ...Getting Started Guide Keysight M8920A PXIe Radio Test Set...
Страница 8: ...viii...
Страница 38: ...38 M8920A Getting Started Guide...
Страница 66: ...66 M8920A Getting Started Guide...
Страница 74: ...74 M8920A Getting Started Guide...
Страница 78: ...78 M8920A Getting Started Guide...
Страница 86: ...86 M8920A Getting Started Guide...
Страница 89: ...Cable and Connector Care M8920A Getting Started Guide 89 Male connector...