Calibrating and Adjusting
3
2000/3000 X-Series Oscilloscopes Service Guide
61
Figure 10
User Calibration cable for 2-channel oscilloscope
b
For a 4-channel oscilloscope, connect BNC tees to the equal-length cables
as shown below. Then connect a BNC(f)-to-BNC(f) (barrel connector) to the
tee as shown below.
Figure 11
User Calibration cable for 4-channel oscilloscope
4
Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on
the rear panel to the BNC barrel connector.
5
Press the
[Utility]
key; then, press the
Service
softkey.
6
Begin the Self Cal by pressing the
Start User Cal
softkey.
Longer cable
to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4