56
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Dual Cursor Accuracy
Range
Power Supply Setting
Test Limits
Channel 1 Channel 2 Channel 3* Channel 4*
5 V/Div
35 V
34.0 V to 36.0 V
________
________
________
________
2 V/Div
14 V
13.6 V to 14.4 V
________
________
________
________
1 V/Div
7 V
6.8 V to 7.2 V
________
________
________
________
500 mV/Div
3.5 V
3.4 V to 3.6 V
________
________
________
________
200 mV/Div
1.4 V
1.36 V to 1.44 V
________
________
________
________
100 mV/Div
700 mV
680 mV to 720 mV
________
________
________
________
50 mV/Div
350 mV
340 mV to 360 mV
________
________
________
________
20 mV/Div
140 mV
136 mV to 144 mV
________
________
________
________
10 mV/Div
70 mV
68 mV to 72 mV
________
________
________
________
5 mV/Div
35 mV
34 mV to 36 mV
________
________
________
________
2 mV/Div
14 mV
13.2 mV to 14.8 mV
________
________
________
________
1 mV/Div
7 mV
6.2 mV to 7.8 mV
________
________
________
________
Band wid th (-3 dB)
Model
Test Limits
Channel 1 Channel 2 Channel 3* Channel 4*
310x
-3 dB at 1 GHz
________
________
________
________
305x
-3 dB at 500 MHz
________
________
________
________
303x
-3 dB at 350 MHz
________
________
________
________
302x
-3 dB at 200 MHz
________
________
________
________
301x
-3 dB at 100 MHz
________
________
________
________
Time Base Accuracy Limits
Calculated
time base
accuracy
limit (ppm)
Measured
time base
error (ppm)
Pass/Fail
Time Base Accuracy Limit: 25 ppm ±5 ppm per year
(aging)
_________ ________ ________
Internal Trigger Sensitivity
Generator Setting
Test Limits
Channel 1 Channel 2 Channel 3* Channel 4*
1 GHz models:
500 MHz models:
350 MHz models:
200 MHz models:
100 MHz models:
1 GHz
350 MHz
200 MHz
100 MHz
< 10 mV/div: greater of 1 div
or 5 mVpp
________
________
________
_______
>= 10 mV/div: 0.6 div
________
________
________
_______
External Trigger Sensitivity
Generator Setting
Test Limits
Ext Trig In
200 MHz
350 mV
________
100 MHz
200 mV
________
* Where applicable
Содержание InfiniiVision 2000 X Series
Страница 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Страница 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Страница 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...