42
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
2
Set up the power meter.
Set the power meter to display measurements in units of watts.
3
Set up the oscilloscope.
a
Press the
[Default Setup]
key.
b
Set channel 1
Coupling
to
DC
.
c
With 2000 X-Series oscilloscopes, connect a 50 ohm feedthrough
termination.
With 3000 X-Series oscilloscopes, set channel 1
Imped
to
50 Ohm
.
d
Set the time base to 500 ns/div.
e
Set the Volts/Div for channel 1 to 200 mV/div.
f
Press the
[Acquire]
key, then press the
Averaging
softkey.
g
Turn the Entry knob to set
#
Avgs
to 8 averages.
4
Set the signal generator for 1 MHz and six divisions of amplitude.
The signal on the oscilloscope screen should be about five cycles at six
divisions amplitude.
5
Set up the Amplitude measurement
a
Press the
[Meas]
key.
b
Press the
Clear Meas
softkey and then the
Clear All
softkey.
c
Press the
Type:
softkey and use the Entry knob to select
AC RMS - Full Screen
(Std Deviation)
within the select menu.
d
Press the
Add Measurement
softkey.
6
Note the oscilloscope AC RMS - FS(1) reading at the bottom of the screen.
(This is the RMS value with any dc offset removed.)
7
Set the power meter Cal Factor % to the 1 MHz value on the calibration chart
on the power sensor.
8
Note the reading on the power meter and covert to Vrms using the expression:
For example, if the power meter reading is 892 uW, then Vin
1MHz
= (892*10
-6
*
50
Ω
)
1/2
= 211.2 mV
rms
.
Vin
1
MHz
Pmeas
1
MHz
50
Ω
×
=
Содержание InfiniiVision 2000 X Series
Страница 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Страница 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Страница 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Страница 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Страница 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...