Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
33
5
Read the “current” average voltage value as V1.
6
Use the BNC tee and cables to connect the oscilloscope calibrator /power
supply to both the oscilloscope and the multimeter (see
Figure 3
Setting up Equipment for DC Vertical Gain Accuracy Test
7
Adjust the output so that the multimeter reading displays the first Volts/div
calibrator setting value in
(depending on the oscilloscope
model).
8
Disconnect the multimeter.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter