Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
23
To construct the test connector (for use with MSO models only)
Keysight 2000/3000 X-Series oscilloscopes that have digital channels enabled
require the test connector described below. Follow the steps to build the test
connector.
1
Obtain a BNC connector and an 8-by-2 section of Berg strip. A longer strip can
be cut to length using wire cutters.
2
On one side of the Berg strip, solder a jumper wire to all of the pins (shown in
3
On the other side of the Berg strip, solder another jumper wire to all of the pins.
4
Solder the center of the BNC connector to a center pin on one of the rows on
the Berg strip.
5
Solder the ground tab of the BNC connector to a center pin on the other row on
the Berg strip.
Table 3
Materials required to construct the test connectors
Description
Recommended Part
Qty
BNC (f) Connector
Keysight 1250-1032 or
Pomona 4578
1
Berg Strip, 8-by-2
3M .100” x .100” Pin Strip
Header or similar
1 strip, cut to length (8x2)
Jumper wire