Keysight E6953A V2X Test Solution Installation & User Guide
Now, the measured loss is < 0.1 dB which is acceptable for running the V2X test
cases. The test step will pass if these measured values are less than the
configured Tx Power Accuracy (configured as 0.5 dB in this example).
Figure 3-23
Tx Power Accuracy
If needed this test can be re-run across all the frequency range (channels 172
-184) and additional entries can be made into the Cable Loss table (
).