Keysight E1418A Service Guide
63
Adjustment Procedure
Adjustments
Current Adjustment
Current output uses an internal resistance value to calibrate all channels. This
resistance value is approximately 244 Ω.. The current is output adjusted by
measuring the voltage developed across this resistor.
To set the internal resistance value use the following CALibration commands:
CALibration:CONFigure:RESistance?
CALibration:VALue:RESistance <value>
Figure 5-4
Optional Voltage Adjustment Connections
You may also, optionally, use an external resistance value (or
load) on each channel for current output and calibration. If you
wish to specify an external resistor value, you must send the
DIAG:CAL:OUTP CHAN command and perform the adjustment at
each channel output. Any load value may be used but values in
the range of 50 Ω to 500 Ω are recommended.
Note that during the adjustment process self-heating of the
external load may affect the adjustment.
Содержание E1418A Option 002
Страница 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...6 ...
Страница 10: ...10 Keysight E1418A Service Guide ...
Страница 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Страница 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Страница 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Страница 87: ...Keysight E1418A Service Guide 81 Repair Maintenance Guidelines Service Figure 6 1 8 Channel Disassembly ...
Страница 88: ...82 Keysight E1418A Service Guide Service Repair Maintenance Guidelines Figure 6 2 16 Channel Disassembly ...
Страница 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Страница 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Страница 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Страница 105: ......