Keysight E1418A Service Guide
Verification Tests
Performance Verification Tests
Performance Verification Tests
These procedures test the electrical performance of the Keysight E1418A using
the specification in Appendix A as the performance standards. These tests are
suitable for incoming inspections, troubleshooting, and preventive measures.
DC Voltage Output Verification
This procedure tests the dc voltage output performance of the module.
1
For each channel in the module, connect the 10 kΩ resistor and DMM as
shown in Figure 4-1. Channel 1 connections are shown in the figure. Connect
the other channels in a similar manner. Set the DMM to measure dc volts.
2
Set the channel to output each of the following voltages:
+16.00
– 4.00
+12.00
– 8.00
+ 8.00
–12.00
+ 4.00
–16.00
0.00
Figure 4-1
DC Voltage Output Connections (Channel 1)
Содержание E1418A Option 002
Страница 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...6 ...
Страница 10: ...10 Keysight E1418A Service Guide ...
Страница 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Страница 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Страница 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Страница 87: ...Keysight E1418A Service Guide 81 Repair Maintenance Guidelines Service Figure 6 1 8 Channel Disassembly ...
Страница 88: ...82 Keysight E1418A Service Guide Service Repair Maintenance Guidelines Figure 6 2 16 Channel Disassembly ...
Страница 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Страница 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Страница 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Страница 105: ......