Keysight E1418A Service Guide
65
Adjustment Procedure
Adjustments
Current output adjustment uses two of the CALibration subsystem queries.
Current adjustment requires voltage measurements.
CALibrationn:CONFigure:CURRent?
CALibrationn:VALue:CURRent? <value>
Once the calibration resistance value is set as described earlier, the general
procedure for each channel current output adjustment is:
1
Make the connections shown in figure 5-6 or 5-7.
2
Send the CALn:CONF:CURR? query to the module and read the response until
a 1 is returned (typically the first query response).
3
Read the voltage on the DMM.
Figure 5-5
Internal Resistor Adjustment Connections
The current adjustment procedure measures and inputs a voltage
value developed across a known resistor value (set at the start of
the procedure).
Содержание E1418A Option 002
Страница 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...6 ...
Страница 10: ...10 Keysight E1418A Service Guide ...
Страница 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Страница 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Страница 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Страница 87: ...Keysight E1418A Service Guide 81 Repair Maintenance Guidelines Service Figure 6 1 8 Channel Disassembly ...
Страница 88: ...82 Keysight E1418A Service Guide Service Repair Maintenance Guidelines Figure 6 2 16 Channel Disassembly ...
Страница 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Страница 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Страница 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Страница 105: ......