Keysight E1418A Service Guide
59
Making Connections
Adjustments
Making Connections
All adjustments can be performed using the CAL output terminals. The CAL
output terminals provide one set of connections for both voltage and current
adjustment. Procedures in this chapter assume connections are made at the CAL
output terminals on the screw-type terminal module.
For demanding applications or for adjustment using the intended
load, you can perform the adjustment at each channel output (to
include the output relay contacts, wiring, and the load in the
path). To make the adjustments at each channel output, use the
DIAG:CAL:OUTP CHAN command to set the adjustment point at
the channels instead of the CAL output terminals. Adjustment at
each channel output will require individual four-wire connections.
Figure 5-1
CAL Output Terminals
Содержание E1418A Option 002
Страница 1: ...Keysight E1418A 8 16 Channel D A Converter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...6 ...
Страница 10: ...10 Keysight E1418A Service Guide ...
Страница 32: ...26 Keysight E1418A Service Guide Operating Information Operator s Checks ...
Страница 77: ...Keysight E1418A Service Guide 71 Adjustment Procedure Adjustments ...
Страница 78: ...72 Keysight E1418A Service Guide Adjustments Adjustment Procedure ...
Страница 87: ...Keysight E1418A Service Guide 81 Repair Maintenance Guidelines Service Figure 6 1 8 Channel Disassembly ...
Страница 88: ...82 Keysight E1418A Service Guide Service Repair Maintenance Guidelines Figure 6 2 16 Channel Disassembly ...
Страница 90: ...84 Keysight E1418A Service Guide Service Repair Maintenance Guidelines ...
Страница 100: ...94 Keysight E1418A Service Guide Replaceable Parts Ordering Information ...
Страница 102: ...96 Keysight E1418A Service Guide Manual Changes Introduction ...
Страница 105: ......