20
WMEMory Subsystem
498
Keysight 8990B Programming Guide
Overview
The
WMEMory
command subsystem allows you to configure the display, offset, and
scale of the waveform in a 8990B memory slot as well as save a waveform into the
selected memory slot. This command subsystem also enables you to clear a
waveform memory slot or query the waveform unit of the selected memory slot.
In this command subsystem, the
{1|2|3|4}
following
WMEMory
represents the
respective memory slot of the 8990B.
Keyword
Parameter form
Note
Page
WMEMory{1|2|3|4}
:CLEar
:DISPlay
<boolean>
:OFFSet
<numeric_value>
:SAVE
<character_data>
:SCALe
<numeric_value>
:UNIT?
[query only]
Содержание 8990B
Страница 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Страница 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 46: ...1 8990B Remote Operation 46 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 142: ...4 ACQuire Subsystem 142 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 156: ...5 CALibration Subsystem 156 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 188: ...6 CHANnel Subsystem 188 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 450: ...15 STATus Subsystem 450 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 460: ...16 DISPlay Subsystem 460 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 508: ...20 WMEMory Subsystem 508 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 518: ...21 SERVice Subsystem 518 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...