22 Keysight 8478B Thermistor Mount
Operation
Calibration Data Application
2-
correction factor when a tuner is used to match the thermistor mount to the
transmission line or RF source. In this case, all of the RF power incident upon
the mount is absorbed in the mount. Since all power is absorbed in the mount,
measurement uncertainty due to mount SWR is eliminated; however, losses in
the tuner must be considered.
Calibration Data Application
When the 8478B is used with the Model 431 or 432 Power Meters, Calibration
Factor or Effective Efficiency corrections can be made by setting a front panel
switch. With the proper setting, the 431 or 432 compensates for the
Calibration Factor or Effective Efficiency in the 8478B. If the 8478B is used with
a power meter other than the 431 or 432, Calibration Factor or Effective
Efficiency corrections can be made by dividing the measured power by the
Calibration Factor or Effective Efficiency value respectively.