44
Keysight 41800A Active Probe
Performance Tests
1 dB Gain Compression Tests
4-
8. Disconnect N(m)-N(m) Cable from the power splitter, and connect the
41800A between the power splitter and N(m)-N(m) Cable, as shown in
(2).
9. Press the Keysight 4195A's
TRIG/RESET
key to make a single sweep
measurement.
10.Enter the Keysight 4195A “A=A-C” command using the following key
sequence.
CLR LINE,
blue shiftkey
, A, =, A, -, C, ENTER/EXECUTE
11.Enter the Keysight 4195A “A=A-A(1)” command using the following key
sequence.
CLR LINE,
blue shiftkey
, A, =, A, -, A, MATH OPERATOR,
'(', 1, ')', ENTER/EXECUTE
12.Press the
SCALE REF
key and the
'A AUTO SCALE'
softkey for auto scaling.
13.Move the o marker to the point at which the
MAG
(data A) displayed in the
upper right corner of the Keysight 4195A's display indicates approximately
–1 dBm.
14.Sequentially press the Keysight 4195A's
MENU
key, and the
'TRIGGER
menu'
and
'MANUAL mode'
softkeys.
15.Conform that the power meter reading is greater than 3 dBm.