36
Keysight 41800A Active Probe
Performance Tests
Calibration Cycle
4-
Calibration Cycle
The 41800A requires periodic performance verification. The 41800A should be
checked out using the performance test at least once a year or more
depending on its frequency of use. Preventive maintenance should be
performed at least twice a year to keep down-time to a minimum, and to insure
optimum operation.
Preparation
This paragraph provides the information which you should know and the steps
that you should perform before starting the performance test.
1. The test equipment must be allowed to warm-up and stabilize for at least
30 minutes.
2. A Keysight 4195A is required to test the 41800A. In the remainder of this
manual, softkeys are indicated in boldface type and enclosed in single
quotes (e.g.,
'NETWORK'
softkey), and keys are indicated in boldface type
only (e.g.
PRESET
key).
Probe Adapter
Keysight 11880A
1
No substitute
2
P,A,T
Cable
Keysight 11500B
2
N(m)-N(m), 50
Ω
, 61 cm
P,A
Adapter
1
1
N(f)-N(f), 50
Ω
N(m)-N(m), 50
Ω
P,A
P,A
Active Probe
Keysight 41800A
1
Frequency: 50 MHz
T
1. P, A, and T represents the Performance Test, Adjustment, and Troubleshooting,
respectively.
2. Used with N(m)-BNC(f) adapter and probe-BNC(m) adapter
Table 4-1
Recommended Test Equipment (Continued)
Equipment
Recommended Model
Qty. Requirements
Use
1