Keysight 41800A Active Probe
Operation Note
4 Performance
Tests
Introduction
This section provides performance test procedures to ensure that the 41800A
meets the specifications listed in
. The performance
test can be performed without accessing the interior of the 41800A.
The test results should be recorded into the Performance Test Record, which is
located at the end of this section.
Equipment Required
The equipment required for performance testing is listed in
Substitutions can be made if the substitution equipment meets or exceeds the
specifications listed in the Requirements column.
The recommended models 4195A and 11880A are obsolete. The
performance and adjustment test shownn in this chapter are to be used as
references only.
Table 4-1
Recommended Test Equipment
Equipment
Recommended Model
Qty. Requirements
Use
1
Network Analyzer
Keysight 4195A
1
No substitute
P,A,T
Power Meter
Keysight 436A or 438A
1
Accuracy:
~
0.1 dB
Frequency: 500 MHz
P
Power Sensor
Keysight 8482A
1
Frequency: 500 MHz
Compatible with the Keysight 436A or 438A P
Power Splitter
Keysight 11667A
1
Frequency Range: 5 Hz to 500 MHz
P,A