Performance Tests
Probe Gain/Frequency Response Tests
(Confirm that the table sweep parameter is frequency. If it is not frequency,
press the
'SWP select'
softkey repeatedly until it changes to frequency.)
'X REG dump'
(move the cursor to the N = 96 position by pressing the down arrow key)
CLR LINE, 2, 0, 0, MHz/V, ENTER/EXECUTE
(move the cursor to the N = 88 position)
CLR LINE, 5, 0, MHz/V, ENTER/EXECUTE
(move the cursor to the N = 1 O position)
CLR LINE, 5, 0, Hz/dBµV, ENTER/EXECUTE, 'set end',
'PROG SWP on off'
(
“on”
in this softkey will change to
intensified green.)
4. Sequentially press the following Keysight 4195A front panel keys to perform
a normalize (thorough) calibration, and wait until the message “Cal
completed” is displayed.
CAL, 'TRANS CAL menu'' 'NORMLIZE (THRU)' I 'THRU'' ENTER/EXECUTE
5. Press the Keysight 4195A's
CAL
key and the
'CORRECTN on off'
softkey to
activate the Keysight 4195A's correction capability. The “
on
” in this softkey
will change to intensified green.
(1)) from the power splitter, and
connect the 41800A between the power splitter and CABLE 1 as shown in
(2).
<<Probe Gain Test>>
7. Press the Keysight 4195A's
TRIG/RESET
key to make a single sweep
measurement.
8. Press the
SCALE REF
key and the
'A AUTO SCALE'
softkey for auto scaling.
9. Move the o marker on the Keysight 4195A's display to the 50 MHz sweep
point by rotating the rotary knob on the Keysight 4195A's front panel.
10.Confirm that the T/R data displayed in the upper right corner of the
Keysight 4195A's display is 0 dB ± 0.5 dB.
<<Frequency Response>>
11.Enter the Keysight 4195A “A=A-A(88)” command using the following key
stroke sequence to have the Keysight 4195A display the frequency response
test results.
CLR LINE,
blue shiftkey
, A, =, A, -, A, MATH OPERATOR,
'(', 8, 8, ')', ENTER/EXECUTE