Keysight 41800A Active Probe
23
Operation
Preparation for Use
Equipment
• Spectrum Analyzer
Any compatible
• Probe-BNC(m) Adapter
As shipped
• Adapters
As necessary
Procedure
1. Connect the equipment as shown in
2. Set up the spectrum analyzer as follows.
3. Confirm that the signal level is
(calibration output level + 6 dB) ± 5 dB
.
10:1/100:1 Divider Adjustments
This adjustment procedure is used to adjust the 10:1/100:1 divider's gain. This
adjustment requires the Keysight 4195A, any compatible Network Analyzer, or
any compatible Spectrum Analyzer.
Figure 3-6
Divider Adjustment Setup
If the spectrum analyzer has a tracking generator output, use the tracking
generator output connector instead of the calibration output connector.
Then set the output level to
~
–10 dBm.
Center Frequency
Same as Calibration Output frequency
Span
0 Hz
Attenuation
40 dBm