KeysightE1406A Command Module Service Guide
97
Introduction
Error Messages
52
DEVICE OUTSIDE OF LADD
WINDOW
A module in an (extender) mainframe is outside of the logical address
window set by the resource manager or set by the
user-defined extender table
.
Either reset the logical addresses or
download a new extender table
.
53
INVALID EXTENDER A24
WINDOW
The resource manager found an invalid start address or size for
an extender A24 address window
.
Either reconfigure the VME
memory devices or use the extender table.
54
DEVICE OUTSIDE OF A24
WINDOW
A module with A24 memory is located outside of the extender
logical address window
.
Either reconfigure the VME memory
devices or use the extender table.
55
INVALID EXTENDER A32
WINDOW
The resource manager found an invalid start address or size for
an extender A32 address window
.
Either reconfigure the VME
memory devices or use the extender table.
56
DEVICE OUTSIDE OF A32
WINDOW
A module with A32 memory is located outside of the extender
logical address window
.
Either reconfigure the VME memory
devices or use the extender table.
57
INVALID UDEF LADD WINDOW
A user-defined logical address window violates the VXI
-
6
specification (has an invalid base or size)
.
58
INVALID UDEF A16 WINDOW
A user-defined A16 window violates the VXI
-
6 specification (has
an invalid base or size)
.
59
INVALID UDEF A24 WINDOW
A user-defined A24 window violates the VXI
-
6 specification (has
an invalid base or size)
.
60
INVALID UDEF A32 WINDOW
A user-defined A32 window violates the VXI
-
6 specification (has
an invalid base or size)
.
61
INVALID UDEF EXT TABLE
The valid flag in the user-defined extender table is not true (1)
.
The valid flag must be set to ’1’ or the table is assumed to be
invalid
.
To disable the table without re
-
booting
,
set the table
address to ’
0
’ using VXI:CONF:ETAB
0.
62
INVALID UDEF EXTENDER TABLE
DATA
There is an invalid number of records in the user-defined
extender table
.
The number of records must be a number
between 1 and 254.
63
UNSUPPORTED UDEF TTL TRIGGER
There is a user-defined extender table TTL trigger entry for a VXI-MXI
extender that does not support TTL triggers.
64
UNSUPPORTED UDEF ECL TRIGGER
There is a user-defined extender table ECL trigger entry for a VXI-MXI
extender that does not support ECL triggers.
65
DEVICE NOT IN CONFIGURE STATE
A message
-
based device was not in the CONFIGURE state
during a re
-
boot
.
The * SYSRESET should propagate to all
mainframes through the INTX cables. Check the INTX
connectors on remote mainframes.
Table 5-3
Command Module Resource Manager Configuration Errors (continued)
Содержание VXI bus 75000 C Series
Страница 1: ...Keysight E1406A Command Module Service Guide 75000 Series C...
Страница 2: ......
Страница 6: ...6...
Страница 10: ...x Keysight E1406A Command Module Service Guide...
Страница 70: ...70 Keysight Keysight E1406A Command Module Service Guide Replaceable Parts Component Locators...
Страница 77: ...Keysight E1406A Command Module Service Guide 77 Assembly Disassembly Instructions Service...
Страница 78: ...78 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 79: ...Keysight E1406A Command Module Service Guide 79 Assembly Disassembly Instructions Service...
Страница 81: ...Keysight E1406A Command Module Service Guide 81 Assembly Disassembly Instructions Service...
Страница 82: ...82 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 83: ...Keysight E1406A Command Module Service Guide 83 Assembly Disassembly Instructions Service...
Страница 84: ...84 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 106: ...106 KeysightE1406A Command Module Service Guide Error Messages Introduction...
Страница 129: ......