KeysightE1406A Command Module Service Guide
93
Introduction
Error Messages
6
A32 MEMORY OVERFLOW
There is not enough available A32 memory required for a device. The
allowable memory space is from 20000000h to FFFFFFFFh (E0000000h -
FFFFFFFFh is only used if there is a 2000 Mbyte device in the system). If
your system has (mainframe) extenders, try using the user-defined
extender table to allocate the memory more efficiently.
7
DC DEVICE MOVE FAILED
A dynamically configured device did not move to its new logical address.
After setting a DC device (or a block of devices), the resource manager
checks the new address(es) to see if the device actually moved.
8
INACCESSIBLE A24 MEMORY
An A24 memory device has memory below 200000h or above DFFFFFh.
The command module cannot access this memory.
9
UNABLE TO MOVE DC DEVICE
There is no logical address (or address block) available for a dynamically
configured device to move to. Try using a user-defined dynamic
configuration table or the user-defined extender table to assign the
addresses more efficiently.
1
0
INSUFFICIENT SYSTEM MEMORY
Too many instruments installed for the amount of RAM available in the
command module. Only the system instrument is started.
11
INVALID INSTRUMENT ADDRESS
A module’s logical address is not a multiple of 8, or is not part of a virtual
instrument. Secondary GPIB addresses are only given to devices with
logical addresses that are a multiple of eight.
12
INVALID UDEF COMMANDER LADD
The user-defined commander logical address is not a valid
commander
.
Either the commander does not exist
,
or it is not a
message
-
based commander
.
14
INVALID UDEF SECONDARY
ADDRESS
Invalid user-defined secondary address specified in the
commander/servant hierarchy table
.
The secondary address
specified was not
0
-
30
,
the address was
0
which is the
command module address, or the module in not in the servant
area of the command module
.
15
DUPLICATE SECONDARY ADDRESS
The same secondary address way specified for more than one module in
the user-defined commander/servant hierarchy table.
16
INVALID SERVANT AREA
The servant area of a commander is greater than 255
,
or the
servant area of a servant module is greater than that of its
commander
.
An invalid servant area is truncated to an allowable
range and system configuration continues.
17
SLOT
0
FUNCTIONS
DISABLED
The command module is installed in slot
0
and its Slot
0
and
System
Controller switches are set to ’Disable’
.
18
INVALID COMMANDER LADD
The commander specified in the user-defined
commander/servant hierarchy table is not a valid
message
-
based commander
,
or the device does not exist
.
Table 5-3
Command Module Resource Manager Configuration Errors (continued)
Содержание VXI bus 75000 C Series
Страница 1: ...Keysight E1406A Command Module Service Guide 75000 Series C...
Страница 2: ......
Страница 6: ...6...
Страница 10: ...x Keysight E1406A Command Module Service Guide...
Страница 70: ...70 Keysight Keysight E1406A Command Module Service Guide Replaceable Parts Component Locators...
Страница 77: ...Keysight E1406A Command Module Service Guide 77 Assembly Disassembly Instructions Service...
Страница 78: ...78 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 79: ...Keysight E1406A Command Module Service Guide 79 Assembly Disassembly Instructions Service...
Страница 81: ...Keysight E1406A Command Module Service Guide 81 Assembly Disassembly Instructions Service...
Страница 82: ...82 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 83: ...Keysight E1406A Command Module Service Guide 83 Assembly Disassembly Instructions Service...
Страница 84: ...84 Keysight E1406A Command Module Service Guide Service Assembly Disassembly Instructions...
Страница 106: ...106 KeysightE1406A Command Module Service Guide Error Messages Introduction...
Страница 129: ......