background image

U3020AS62

Instrument Settings

GPIB Address Command

GPIB Address Command Description
This command assigns a GPIB address to the U3020A. 

Syntax

SYSTem:COMMunicate:GPIB:ADDRess <address> SYSTem:COMMunicate:GPIB:ADDRess?

If you change the GPIB address, you must cycle power on the U3020A to 

activate the new address.

Example

The following command sets the GPIB address to 10: SYST:COMM:GPIB:ADDR 10
The following query returns the current GPIB address: SYST:COMM:GPIB:ADDR?
Typical Response: 10

For a complete list of SCPI commands, refer to the Keysight L449xA 

RF Switch Platform User’s Guide (L4490-90001).

Keysight U3020AS62 User's and Service Guide

21

Содержание U3020AS62

Страница 1: ...Keysight Technologies U3020AS62 DC to 26 5 GHz Signal Distribution Test Set User s and Service Guide...

Страница 2: ...perform display or disclose commercial computer software or commercial computer software documentation No additional government requirements beyond those set forth in the EULA shall apply except to t...

Страница 3: ...AN Interface 15 Locating the Instrument 15 Configuring the GPIB Interface 16 Adding Instruments to the GPIB Configuration 17 Changing the GPIB Address 17 IP Address Command 17 Auto IP Address Command...

Страница 4: ...nnector Care and Cleaning Precautions 32 Safety and Regulatory Information 33 Instrument Markings 33 EMC Compliance 34 South Korean Class A EMC Declaration 34 Safety 34 Acoustic Statement European Mac...

Страница 5: ...U3020AS62 Keysight U3020AS62 User s and Service Guide 5...

Страница 6: ...r If the container or packing material is damaged it should be kept until the contents of the shipment have been checked mechanically and electrically If there is physical damage refer to Contacting K...

Страница 7: ...degrees C This product has an auto ranging line voltage input Be sure the supply voltage is within the specified range Keysight U3020AS62 User s and Service Guide 7 Definitions Specifications describe...

Страница 8: ...or table top of adequate size and strength General RF Switch Specifications Table 2 Test Set Dimensions Item Length Weight Width 42 5 cm 16 73 in Depth 58 cm 22 83 in Weight 8 kg 17 64 lb Table 3 Int...

Страница 9: ...ing dB Phase Tracking Deg DC to 1 0 GHz 0 05 0 7 1 0 to 6 0 GHz 0 1 4 0 6 0 to 12 GHz 0 2 8 0 12 to 20 GHz 0 3 13 0 20 to 26 5 GHz 0 8 18 0 Propagation Delay Channel A to B Frequency Range Skew ps 0 1...

Страница 10: ...power on sequence which requires several seconds to complete The LEDs provide information on the state of the instrument during power on and during upgrades of the instrument firmware Table 6 on page...

Страница 11: ...due to disconnected LAN cable failure to acquire and IP address waiting for DHCP assigned address ATTN LAN PWR Green flashing Green Green Instrument Busy State firmware download LAN LED red if downlo...

Страница 12: ...he operator The detachable power cord is the instrument disconnecting device It disconnects the mains circuits from the mains supply before other parts of the instrument The front panel switch is only...

Страница 13: ...ive instrument LAN line Open the internet browser application on the computer Enter this default web address on your browser application http U3020aS62 xxxx The Welcome screen for the instrument shoul...

Страница 14: ...ls A and B All other remaining Port Inputs will be terminated by an internal 50 Ohm load ROUT SEQ TRIG SEL_PORT1 Table 7 Test Set Control Commands Commands Description Comments SEL_PORT1 Select Port 1...

Страница 15: ...the application window Figure 4 Keysight Connection Expert Locating the Instrument Keysight Connection Expert opens with a Welcome Screen and a window similar to that shown in Figure 4 The available...

Страница 16: ...ments are added to the configured interface Figure 6 Adding Instruments Configuring the GPIB Interface Programming access to the Test Set is also available through the instrument s GPIB interface The...

Страница 17: ...ck Change Properties in the Configurable Properties application window change the address to the new address setting and select OK IP Address Command This command assigns a static Internet Protocol IP...

Страница 18: ...ration Protocol DHCP Syntax SYSTem COMMunicate LAN DHCP OFF 0 ON 1 SYSTem COMMunicate LAN DHCP When DHCP is enable factory setting the instrument will try to obtain an IP address from the DHCP server...

Страница 19: ...he DNS address currently being used by the instrument the quotes are also returned SYST COMM LAN DNS Typical Response 198 105 232 4 Domain Name Command This command assigns a Domain Name to the U3020A...

Страница 20: ...dress SYST COMM LAN GATEWAY 255 255 20 11 The following query returns the Default Gateway address currently being used by the instrument the quotes are also returned SYST COMM LAN GATEWAY Typical Resp...

Страница 21: ...change the GPIB address you must cycle power on the U3020A to activate the new address Example The following command sets the GPIB address to 10 SYST COMM GPIB ADDR 10 The following query returns the...

Страница 22: ...uipment Setup 1 Configure PNA and Test Set as shown in Figure 8 2 Turn On the PNA and Test Set The Test Set LED Indicators should indicate the following after 1 minute ATTN OFF LAN ON PWR ON 3 Preset...

Страница 23: ...T1 to select Port 1 Figure 9 is representation of the S21 response trace Markers shown are not provided in the procedure above Refer to Table 4 on page 9 for typical Insertion Loss values 8 To verify...

Страница 24: ...rge Protection Electrostatic discharge ESD can damage or destroy electronic components The instrument is shipped in materials that prevent damage from static and should only be removed from the packag...

Страница 25: ...interconnects of the Test Set The main function of this Test Set is Dual Channel Signal Distribution The Test Set consists of six main components a controller module a 39495EXT module distribution bo...

Страница 26: ...dles all the communication between the Controller and the Test Set via LAN or GPIB connectivity 39495EXT Module This module provides the bias and control signals for the RF Switches Distribution Board...

Страница 27: ...ing the front panel button continue with the following procedure a Remove the top cover of the instrument b Ensure that the ribbon cable and 2 wire connector are properly in place c Measure the DC vol...

Страница 28: ...resent suspect the front panel Input Port LED board or the Controller Module I O section RF Switch is Not Switching 1 If the RF Switches are not operating continue with the following procedure a Deter...

Страница 29: ...11 Top View Controller Module and Power Supply Distribution Bd Y1151A Switch Driver 34945 Ext RF Switches Channel A and B I O Distribution Bd Digital I O Relay Driver L4490A Option 004 B Keysight U30...

Страница 30: ...tact Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous Intentional interruption is prohibited Always use the three prong AC power co...

Страница 31: ...ay cause product damage and serious injury This product is designed for use in Installation Category II and Pollution Degree 2 Before switching on this instrument make sure the supply voltage is in th...

Страница 32: ...ers The detachable power cord is the instrument disconnecting device It disconnects the mains circuits from the mains supply before other parts of the instrument The front panel switch is only a stand...

Страница 33: ...e design was proven It indicates that the product complies with all relevant directives The Keysight email address is required by EU directives applicable to our product The CSA mark is a registered t...

Страница 34: ...opean Machinery Directive Acoustic noise emission LpA 70 dB Operator position Normal operation mode per ISO 7779 To find a current Declaration of Conformity for a specific Keysight product go to https...

Страница 35: ...ne the warranty status of your unit Shipping Your Product to Keysight for Service or Repair IMPORTANT Keysight Technologies reserves the right to reformat or replace the internal hard disk drive in yo...

Страница 36: ...This information is subject to change without notice Keysight Technologies 2013 2020 Edition 3 October 2020 Supersedes June 2020 U3020 90011 www keysight com...