U3020AS62
Instrument Settings
GPIB Address Command
GPIB Address Command Description
This command assigns a GPIB address to the U3020A.
Syntax
SYSTem:COMMunicate:GPIB:ADDRess <address> SYSTem:COMMunicate:GPIB:ADDRess?
If you change the GPIB address, you must cycle power on the U3020A to
activate the new address.
Example
The following command sets the GPIB address to 10: SYST:COMM:GPIB:ADDR 10
The following query returns the current GPIB address: SYST:COMM:GPIB:ADDR?
Typical Response: 10
For a complete list of SCPI commands, refer to the Keysight L449xA
RF Switch Platform User’s Guide (L4490-90001).
Keysight U3020AS62 User's and Service Guide
21