U3020AS62
Controlling the Test Set
Test Set Control Commands
The Test Set has predefine sequence command sets defined at the factory and stored in the
non-volatile memory of the controller module. If the specified sequence name is not currently
stored in the memory, due to corrupted program or accidentally deleted, an error will be
generated. Refer to the Service section of this manual for instructions to remedy this issue.
Syntax
ROUTe:SEQuence:TRIGger <command>
Parameters
Refer to
for commands use to control the Test Set’s programmable switches.
Example
The following executed sequence command, "SEL_PORT1", will position the RF switches to
create a signal path through Port 1, Input to Output for Channels A and B. All other remaining
Port Inputs will be terminated by an internal 50 Ohm load.
ROUT:SEQ:TRIG SEL_PORT1
Table
7
Test Set Control Commands
Commands
Description
Comments
SEL_PORT1
Select Port 1 Input of Chan A and B
All unselected Port Inputs are terminated
SEL_PORT2
Select Port 2 Input of Chan A and B
All unselected Port Inputs are terminated
SEL_PORT3
Select Port 3 Input of Chan A and B
All unselected Port Inputs are terminated
SEL_PORT4
Select Port 4 Input of Chan A and B
All unselected Port Inputs are terminated
SEL_PORT5
Select Port 5 Input of Chan A and B
All unselected Port Inputs are terminated
SEL_PORT6
Select Port 6 Input of Chan A and B
All unselected Port Inputs are terminated
TERM_ALL_INPUTS
Terminate all Input Ports (50 Ohm)
Output Ports are not terminated
LED_ALL
All LEDs ON
Diagnostic Check
LED_OFF
All LEDs OFF
Diagnostic Check
14
Keysight U3020AS62 User's and Service Guide