3 Review Hardware Components

AC Line Input

(Use the AC line cord supplied with the S9100A.)

Connector, Three-Prong

100/120 V, 50/60 Hz,
1200 W MAX (Lower range)
220/240 V, 50/60 Hz,
1300 W MAX (Upper range)

See also

AC Power Cord

 on page 

14

.

POWER SYNC
(DO NOT CONNECT TO LAN!)

Connector, IN and OUT

RJ-45 cables (CAT5 or better not exceeding
two meters in length) can be used to connect
multiple PXI chassis; up to four chassis may
be connected together for power-up syn-
chronization. DO NOT connect to the LAN!

When connected with RJ-45 cables, the
power button on any chassis may be used to
power up or power down the entire system.

FAN and INHIBIT Switches
(Used for remote inhibit and power rail monitoring.)

FAN Switch

INHIBIT Switch

The FAN switch controls the fan speed:
HIGH - the fan voltage duty cycle is set
to 100%, which generates the highest
fan speeds and best chassis cooling
(as well as the most fan noise).

AUTO - the speed of the three fans are
controlled based on the temperature of
the chassis and the setting on the
Soft Front Panel (SFP) Temperature
parameter. See the

Keysight PXIe Chassis

User Guide (M9019-90003)

.

The INHIBIT switch controls the method
of powering up the chassis:
DEF (default) - recommended position
when connecting AC power cord.
Front panel power button is used to
switch between ON and Standby and
requires that an M9024A or M9037A is
installed in system controller slot 1.

MAN (manual) - Inhibit signal on the rear
panel DB-9 connector controls chassis
power. When set to MAN, POWER SYNC
feature is disabled.

26

S9100A 5G Multi-Band Vector Transceiver, Startup Guide

Содержание S9100A

Страница 1: ...Startup Guide S9100A 5G Multi Band Vector Transceiver...

Страница 2: ...ily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy...

Страница 3: ...t Panel 21 3 2 2 Connectors RF Transceiver on the Rugged Front Panel 22 3 2 3 Connectors Head 1 on the Rugged Front Panel 22 3 2 4 Connectors M9037A PXIe Embedded Controller 23 3 2 5 Connectors S9100A...

Страница 4: ...ection String Examples for S9100A 022 69 7 3 1 4 Using Keysight Connection Expert to Retrieve VISA Addresses 70 7 3 2 Using X Apps to Perform Alignments 74 7 3 3 Using the Power Calibration Tab 75 7 3...

Страница 5: ...and millimeter wave FR2 24 25 to 43 5 GHz frequencies In this document This document describes the installation process to start up and prepare the Keysight S9100A 5G Multi Band Vector Transceiver fo...

Страница 6: ...2 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 7: ...on page 15 Front and Rear Panel Symbols on page 16 Returning for Service on page 18 2 1 Warning Statements and Symbols The following notices are used throughout this document Familiarize yourself wit...

Страница 8: ...tion This documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain products in a safe condition 2 2 1 Safety Compliance These products...

Страница 9: ...bited The Keysight S9100A 5G Multi Band Vector Transceiver is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 Second Edition and 664 respectively The Keysight S9100A...

Страница 10: ...dles Depth 424 mm 16 7 in across front connectors and rear feet Weight 18 1 kg 40 lbs Keysight M1740A mmWave Transceiver Height 66 mm 2 6 in Width 139 mm 5 5 in Depth 183 mm 7 2 in Weight 2 2 kg 4 85...

Страница 11: ...cord removed and in a well ventilated area Allow all residual alcohol moisture to evaporate and the fumes to dissipate prior to energizing the instrument Keep isopropyl alcohol away from heat sparks a...

Страница 12: ...is designed for use in the following conditions For indoor use only Altitude up to 6 561 68 ft 2 000 m Operating Temperature 10 to 40 C 5 to 85 non condensing relative humidity Samples of these produc...

Страница 13: ...ICES 001 Cet appareil ISM est conforme a la norme NMB 001 du Canada 2 7 1 South Korean Class A EMC Declaration This equipment has been conformity assessed for use in business environments In a residen...

Страница 14: ...ot exceeded by cabinet installation Keysight recommends forced air convection whenever instrument s are installed in a cabinet and further recommends that the maximum operating temperature of the cabi...

Страница 15: ...es or circuit breakers which are readily identifiable and is easily reached by the operator may be used as disconnecting devices The Keysight S9100A Base System requires a 1U space below when rack mou...

Страница 16: ...770 90001 Failure to ground the test set properly can result in personal injury Before turning on the test set you must connect its protective earth terminals to the protective conductor of the main p...

Страница 17: ...interruption of the protective conductor inside or outside of the product is likely to make the product dangerous Intentional interruption is prohibited The M1740A mmWave Transceiver does not have an...

Страница 18: ...chable power cords are readily identifiable and easily reached by the operator The detachable power cord is the instrument s disconnecting device It disconnects the Mains circuits from the Mains suppl...

Страница 19: ...ore connecting any coaxial cable to a test set connector for the first time each day momentarily short the center and outer conductors of the cable together Personnel should be grounded with a 1 M res...

Страница 20: ...complies with all relevant directives The RCM mark is a registered trademark of the Australian Communications and Media Authority The Keysight email address is required by EU directives applicable to...

Страница 21: ...ts are expected to leak or deteriorate during normal use and generally reflects the expected useful life of the product Universal recycling symbol This symbol indicates compliance with the China stand...

Страница 22: ...l numbers The serial number label is located on the side panel of the module The serial number can also be read from the Soft Front Panel interface after the hardware is installed Description of failu...

Страница 23: ...Components on page 20 Connector Descriptions on page 21 Connectors 10 MHz Ref In Out on the Rugged Front Panel on page 21 Connectors RF Transceiver on the Rugged Front Panel on page 22 Connectors Hea...

Страница 24: ...System with one M9410A PXIe VXT Vector Transceiver one E7770A Common Interface Unit CIU with one Channel Card High IF Slot 1 one Channel Card No DUT IF Slot 2 one LO Distribution Card one LO Card one...

Страница 25: ...the Rugged Front Panel 10 MHz Ref Out Connects from M9300A PXIe Reference 10 MHz Out Connector BNC f Amplitude 9 5 dBm nominal 10 MHz Ref In Connects to M9300A PXIe Reference Ref In and locks to anot...

Страница 26: ...ce InputLevel Range 1 k or 50 nominal 3 3 V to 3 3 V OutputImpedance OutputLevel Range 50 nominal 3 3 V LVTTL RF Transceiver Half Duplex Connector Type N f Frequency Range 380 MHzto 6 GHz Amplitude 0...

Страница 27: ...VI D monitor USB 2 0 and 3 0 Connectors Four USB 2 0 TypeA Connectors Two USB 3 0 LAN 1 and LAN 2 Connectors Two 10 100 1000BASE T RJ 45 GigabitEthernetports LANConnector Top LED Amber LANConnector Bo...

Страница 28: ...coin cell lithium battery This battery powers the clock circuitand retains configuration memory in CMOS RAM whilethesystem is turned off To replacetheCMOS Backup Battery 1 Turn off power to thePXIech...

Страница 29: ...Upper range See also AC Power Cord on page 14 POWER SYNC DO NOT CONNECT TO LAN Connector INand OUT RJ 45 cables CAT5 or better notexceeding two meters in length can beused to connect multiplePXI chass...

Страница 30: ...ng FAN Switch INHIBIT Switch The FAN switch controls the fan speed HIGH the fan voltage duty cycle is set to 100 which generates the highest fan speeds and best chassis cooling as well as the most fan...

Страница 31: ...el BNC f 50 nominal Frequency Input InputSignal Inputsignal PXI timing slotPXI_CLK10_IN 10 MHz 100 PPM 100 mVPP to 5 VPP square waveor sine wave 5 V or 3 3 V TTLsignal These 10 MHz REF IN and OUT conn...

Страница 32: ...HzIn If LED is Red LO Card is notlocked to the10 MHzIn 10 MHz In Connector Frequency BNC f 50 nominal 10 MHzreferencesignal input LO Aux Out Connector SMA f 50 nominal LO Card LO Aux Out to LO Distrib...

Страница 33: ...n Card LO Out2 to Channel Card 2 LO In LO Out 3 Connector SMA f 50 nominal LO Distribution Card LO Out3 to Channel Card 3 LO In LO Out 4 Connector SMA f 50 nominal LO Distribution Card LO Out4 to Chan...

Страница 34: ...is used Configurations using a Channel Card equipped with IF circuitry IF connectors do function and also use LO In IF In A Connector Type N f 50 nominal DLIFinputChannel A DUT IF In Out A Connector S...

Страница 35: ...e LO Distribution Card do not extend through to the rear panel there are no rear connections for the LO Card or LO Distribution Card 3 2 6 5 Connectors E7770A CIU Rear Panel Channel Cards The E7770A C...

Страница 36: ...nal useonly CONTROL LAN Connector 10 100 1000BASE T RJ 45 GigabitEthernetPort TRIG 1 Connector BNC f 50 nominal InputLevel Range 5 V TTL Trigger signal below0 8 volts is a zero low and anything above...

Страница 37: ...ined inputsignal to theLO Pwr Ctrl IFIn port LEDs Trig LED A trigger indicator used by Keysightin testing not needed for customer use In normal usage as described for ExtTrig theLED does notlight RF1...

Страница 38: ...e 24 to 44 GHz 24 25 to 29 5 GHzand 37 to 40 GHz 15 VDC 20 dBm Maximum Input RF Tx Rx 2 Connector 2 4 mm f 50 nominal This portcan beconfigured either to supply a mmWavesignal to a DeviceUnder Test DU...

Страница 39: ...r range 2 5 to 4 GHz 20 to 30 dBm minimum CW 24 to 8 dBm CW IF In Connector SMA f 50 nominal This portaccepts an IFinputto theupconverter in theM1740A IFfrequency range IFinputpower range 2 5 to 4 GHz...

Страница 40: ...36 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 41: ...ard in the E7770A Common Interface Unit CIU providing one Tx channel and one Rx channel operating in one frequency range at a time FR1 frequencies ranging from 380 MHz to 6 GHz or FR2 banded mmWave fr...

Страница 42: ...capability to perform Rx tests with a blocking signal interfering signal This configuration can generate a Wanted signal and at the same time generate a Blocking signal a Wanted signal produced from t...

Страница 43: ...4 Install Hardware S9100A 5G Multi Band Vector Transceiver Startup Guide 39 4 2 Keysight S9100A Option RH1 5G Multi Band Vector Transceiver...

Страница 44: ...5 1 S9100 60003 RF Cable SMA m to TNC m 1100 mm 43 3 in Connect E7770A CIU CHANNEL 1A LO CTRL PWR rear panel to S9100A Head 1 Ch 1A In 6 1 S9100 60005 Three Cable Bundle Color Coded Red SMA m to TNC m...

Страница 45: ...i Band Vector Transceiver on page 39 one channel configuration consists of one RF transceiver and one mmWave transceiver providing one Tx channel and one Rx channel operating in one frequency range at...

Страница 46: ...4 Install Hardware 4 3 Keysight S9100A Option 007 mmWave Transceiver with High IF Dual Use Configuration High IF and mmWave 42 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 47: ...L PWR rear panel to S9100A labeled Ch 1A In front panel 6 1 S9100 60005 Three Cable Bundle Color Coded White SMA m to Type N m 4000 mm 157 5 in Connect M1740A IF In Out to S9100A IF In Black SMA m to...

Страница 48: ...e Tx channel and one Rx channel operating in one frequency range at a time FR1 frequencies ranging from 380 MHz to 6 GHz or FR2 banded mmWave frequencies from 24 25 to 43 5 GHz or High IF frequencies...

Страница 49: ...4 Install Hardware S9100A 5G Multi Band Vector Transceiver Startup Guide 45 4 4 Keysight S9100A Option 022 mmWave Transceiver with High IF and Blocker Dual Use Configuration High IF and mmWave...

Страница 50: ...front panel 6 1 S9100 60005 Three Cable Bundle Color Coded White SMA m to Type N m 4000 mm 157 5 in Connect M1740A IF In Out to S9100A IF In Black SMA m to Type N m 4000 mm 157 5 in Connect M1740A IF...

Страница 51: ...o 43 5 GHz or High IF frequencies ranging from 6 to 12 GHz High IF frequencies are provided by using additional High IF input output connectors that route the signal path through up and down converter...

Страница 52: ...48 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 53: ...nd Vector Transceiver Startup Guide 49 5 Install Software If a standard configuration that includes the M9037A PXIe Embedded Controller was ordered all needed software should have come pre installed p...

Страница 54: ...50 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 55: ...ming the following 1 Powering on the Keysight S9100A Base System The M9037A PXIe Embedded Controller and the M9300A PXIe Reference are part of the Keysight S9100A Base System they both power on and in...

Страница 56: ...1250 1250 Adapters SMA f to Type N m 2 If it is not already running start the Modular TRX software and review the interface this interface is referred to as a Mode The Modular TRX interface can be sta...

Страница 57: ...View as Normal d Select OK 5 Set the RF Input to 3 GHz this is the RF In port on the Rugged Front Panel a Select the drop down menu panel top right corner select Input Output select the Input tab sel...

Страница 58: ...ont Panel a Select the drop down menu panel top right corner select Input Output select the RF Source tab select the RF Output Port drop down menu and select RF Output b Set RF Power to 10 dBm Frequen...

Страница 59: ...t the drop down menu panel top right corner and select Peak Search If there is a signal at 3 GHz and approximately 10 dBm the RF Transceiver RF Input and RF Output ports of the S9100A are working prop...

Страница 60: ...GHz this port can be thought of as the RF Input Receiver or Analyzer port on the M1740A mmWave Transceiver a Select the drop down menu panel top right corner and select Input Output select the Input t...

Страница 61: ...t on the M1740A mmWave Transceiver a Select the drop down menu panel top right corner and select Input Output select the RF Source tab select the RF Output Port drop down menu and select Head 1 RFHD 1...

Страница 62: ...el top right corner and select Peak Search If there is a signal at 28 GHz and approximately 10 dBm the RF Tx Rx 1 and RF Tx Rx 2 ports of the S9100A are working properly The power level is only approx...

Страница 63: ...Conditions on page 60 2 Review Calibration Equipment Required on page 61 3 Review Calibration Connectors Ports 1 to 8 on page 62 How to proceed 4 Run S910xA System Calibration Software on page 64 5 C...

Страница 64: ...t two hours before being powered on S9100A system has been powered on continuously for at least two hours warm up time with IQ Analyzer or X Series application e g 5G NR running and the M1740A mmWave...

Страница 65: ...Power Sensor or equivalent to the end of the loop back cable described above that is used during Cable signal path cal ibration It can also be used to connect two 2 4 mm m to m cables during Receiver...

Страница 66: ...HD 1 RRH 1 RFHD 2 M1740A Port 1 2 1 2 RRH 1 RFHD 2 RRH 1 RFHD 1 M1740A Port 2 1 2 1 RF Output RF Input RF Transceiver without a Blocker 3 4 IFIO 1 IFIO 1 High IF E7770A w DUT IF Channel Card 7 8 If ca...

Страница 67: ...In with a Blocker have the same name on the external labeling as ports RF Out and RF In without a Blocker they have different system performance Differences are because S9100A Option 022 have differe...

Страница 68: ...stem Calibration can be started as follows Select the Start menu lower left corner icon Scroll down to K applications Select the Keysight S910xA drop down arrow Scroll down the list and select S910xA...

Страница 69: ...eiver with High IF and Blocker Connection Strings differ depending on whether the Calibration is run on an embedded controller local host or an external controller remote host when run on an embedded...

Страница 70: ...tain the IP address for example TCPIP0 141 121 XXX YYY inst0 INSTR Channel The Channel entry box designates which channel the power sensor is connected to on the power meter On a power meter with two...

Страница 71: ...name of the embedded controller local host Primary Transceiver TCPIP0 localhost hislip0 INSTR Secondary Transceiver VISA Address entry box should remain blank using an external controller remote host...

Страница 72: ...iver TCPIP0 localhost hislip0 INSTR Secondary Transceiver VISA Address entry box should remain blank using an external controller remote host with one M9410A PXIe VXT as the Primary Transceiver the Co...

Страница 73: ...edded controller local host Primary Transceiver TCPIP0 localhost hislip0 INSTR Secondary Transceiver TCPIP0 localhost hislip1 INSTR using an external controller remote host with one M9410A PXIe VXT as...

Страница 74: ...t of applications appear in alphabetical order 2 Select any letter such as A from this list of applications and a selection menu should appear 3 Select the letter K from this selection menu 4 Scroll d...

Страница 75: ...instruments select the M9410A Keysight Technologies localhost b On the right side of the dialog box under Connection Strings highlight the VISA Address to be used on the S910xA System Calibration dia...

Страница 76: ...ss of the Power Sensor and select Ctrl C to copy the VISA Address to the Clipboard this should be of the form USB0 0x2A8D 0xA618 MY59160002 0 INSTR c Open the S910xA System Calibration dialog select t...

Страница 77: ...r Transceiver Startup Guide 73 e Optional Before use Zero Calibrate the Power Sensor by clicking Start Disconnect the power sensor from any RF ports the zero calibration will not be valid if any power...

Страница 78: ...as follows Select the Start menu lower left corner icon Scroll down to K applications Select the Keysight Modular Transceiver drop down arrow Scroll down the list and select LaunchModularTRX 2 Select...

Страница 79: ...Calibrations If the standard configuration is S9100A Option RH1 5G Multi Band Vector Transceiver a Calibration 1 of 2 Power from RRH 1 RFHD 1 RRH 1 RFHD 2 on page 76 b Calibration 2 of 2 Power from R...

Страница 80: ...Power Calibration tab b Select the Operation drop down arrow and select Calibration c Select the Calibration Step s drop down arrow and select Complete d Select the Port Selection drop down arrow and...

Страница 81: ...er 2 4 mm f to 2 4 mm f or equivalent connect the U8487A Power Sensor or equivalent to the end of the cable d Select Ok to continue the calibration View the Results Cable tab when the calibration is c...

Страница 82: ...tab b Select the Operation drop down arrow and select Calibration c Select the Calibration Step s drop down arrow and select Complete d Select the Port Selection drop down arrow and select RRH 1 RFHD...

Страница 83: ...er 2 4 mm f to 2 4 mm f or equivalent connect the U8487A Power Sensor or equivalent to the end of the cable d Select Ok to continue the calibration View the Results Cable tab when the calibration is c...

Страница 84: ...ower Calibration tab b Select the Operation drop down arrow and select Calibration c Select the Calibration Step s drop down arrow and select Complete d Select the Port Selection drop down arrow and s...

Страница 85: ...er 2 4 mm f to 2 4 mm f or equivalent connect the U8487A Power Sensor or equivalent to the end of the cable d Select Ok to continue the calibration View the Results Cable tab when the calibration is c...

Страница 86: ...ab b Select the Operation drop down arrow and select Calibration c Select the Calibration Step s drop down arrow and select Complete d Select the Port Selection drop down arrow and select RRH 1 RFHD 2...

Страница 87: ...er 2 4 mm f to 2 4 mm f or equivalent connect the U8487A Power Sensor or equivalent to the end of the cable d Select Ok to continue the calibration View the Results Cable tab when the calibration is c...

Страница 88: ...e RF Output for M9410A PXIe VXT 2 is set to Off 1 Calibrate the Source Tx signal path a Select the Power Calibration tab b Select the Operation drop down arrow and select Calibration c Select the Cali...

Страница 89: ...uivalent connect one end of the cable to M1740A mmWave Transceiver connector RF Tx Rx 1 c Using an 11900B adapter 2 4 mm f to 2 4 mm f or equivalent connect the U8487A Power Sensor or equivalent to th...

Страница 90: ...alibration View the Results Receiver tab when the calibration is complete Corrections are stored in S9100ID_ SerialNumber Rx RFInputOutput2 CHANNELPOWER csv 86 S9100A 5G Multi Band Vector Transceiver...

Страница 91: ...e RF Output for M9410A PXIe VXT 2 is set to Off 1 Calibrate the Source Tx signal path a Select the Power Calibration tab b Select the Operation drop down arrow and select Calibration c Select the Cali...

Страница 92: ...Power Sensor or equivalent to the end of the cable d Select Ok to continue the calibration View the Results Cable tab when the calibration is complete 3 Calibrate the Receiver Rx signal path a Disconn...

Страница 93: ...7 Run Power Calibration S9100A 5G Multi Band Vector Transceiver Startup Guide 89...

Страница 94: ...90 S9100A 5G Multi Band Vector Transceiver Startup Guide...

Страница 95: ...488 2 Common Commands on page 93 IDN on page 93 RST on page 93 System Subsystem on page 94 SYSTem ERRor NEXT on page 94 SYSTem OPTion on page 94 Status Subsystem on page 94 STATus OPERation ROSCillato...

Страница 96: ...CONFigure PRIMary POWer AMPLitude on page 98 SOURce CONFigure SECondary FREQuency num freq_suffix on page 98 SOURce CONFigure SECondary FREQuency on page 98 SOURce CONFigure SECondary POWer AMPLitude...

Страница 97: ...ver 3 0 184 0 Installer 2 13 5 Fields Manufacturer Keysight Model Number S9100A w E7770A CIU or S9101A PXI CIU Empty string CIU Sever Version string Installer Version string RST The reset command rese...

Страница 98: ...SCillator The status operation query returns the status bit for the register of the reference If the reference module is unlocked this query returns a 1 If the reference module is locked this query re...

Страница 99: ...l ROSCillator SOURce The reference oscillator source command changes the reference frequency source to either use the internal clock or use the front panel external reference Default INTernal ROSCilla...

Страница 100: ...assis trigger route query returns the current routing from front panel trigger 1 or front panel trigger 2 Important By convention the front panel triggers are set on bus segment 2 To forward the trigg...

Страница 101: ...s command exists for maintenance or custom configurations The hislip number of the primary and secondary M941x sources should not need to be changed as they are inferred from the hardware configuratio...

Страница 102: ...cy The source configure secondary frequency command sets the output frequency of the blocking signal The output frequency range is restricted by the port that was set by SOURce CONFigure PORT OUTPut R...

Страница 103: ...ffset command sets the loss between the source output port and the DUT input port This value ends up being set as the power offset on the primary source for the intended signal A loss between the sour...

Страница 104: ...XT Configure settings common to both sources ScpiCommand SOUR CONF PORT SignalPort ScpiCommand SOUR CONF POW OFFS OtaLoss ToString G6 Configure the Primary signal ScpiCommand SOUR CONF PRIM FREQ Prima...

Страница 105: ...S9100A 5G Multi Band Vector Transceiver Startup Guide 101...

Страница 106: ...Transceiver To access the Keysight licensing website http www keysight com find licensing To receive the latest updates by email subscribe to Keysight Email Updates www keysight com find MyKeysight T...

Страница 107: ...This information is subject to change without notice Keysight Technologies 2019 2022 Edition 1 April 2022 S9100 90001 www keysight com...

Отзывы: