50
Keysight N5991 SATA Test Automation Software Platform User Guide
5
Receiver Tests
Parameters in Expert Mode
• Number of Allowed Frame Errors: The number of frame errors that are
allowed. The default value is '0'.
• Test Duration: The duration of the frame error measurement at each SJ
frequency.
Data Generator
• Differential Voltage: The calibrated inner eye height at TP2.
• Random Jitter (RJ): The amount of calibrated RJ added to the signal.
• Sinusoidal Jitter (SJ): The amount of calibrated SJ added to the signal.
• Data Rate Deviation: A fixed deviation from the nominal data rate.
• SSC Deviation: Maximum amount of deviation from the nominal data
rate due to down-spread SSC modulation.
• SSC Frequency: The frequency of the SSC modulation.
Error Detector
• Show Additional Counters: Additional counters, such as the frame
counter, are shown in the result table.
Used Calibrations
• Random Jitter Calibration
• Sinusoidal Jitter Calibration
• Differential Voltage Calibration