Service Information
Test and Adjustment Procedures
should be performed
whenever any active electronic part is replaced in the test set.
Additional procedures are required when certain components are
indicates when additional procedures are required. The
additional procedures depend on the location of the replaced component.
The additional procedures are indicated as A, B, C, D, F, Q, R, S and Z
The procedures referenced above may be found on the following pages. Many
of the steps require familiarity with the
section beginning on
Table 2-4
Additional Required Procedures for Assemblies Replaced
Description of the Replaced
Component
Keysight
Part Number
Location:
Ports 1 & 2
RF Section
1
Location:
Ports 3 & 4
RF Section
Location:
LO Section
General
Location
Millimeter Head Interface Bd
N5261-60101
Z
Test Set Controller Bd
N5261-60006
C
2
D
RF/LO Bias/ALC Bd
N5261-60103
Q
R
S
Preamplifier Bias Bd
N5261-60104
S
RF 4-way Divider
N5262-80003
S
RF Detector
3333-80021
Q
R
S
Coax Cable, Sloped
Attenuator
N5262-20033
Q
R
Preamplifier
5087-7750
S
RF/LO Amplifier
5087-7771
Q A D
R B D
S
RF Directional Coupler
0955-0148
Q A D
R B D
S
RF Switch, SPDT Solid State
5087-7733
Q A D
R B D
1.
Refer to
thru
thru
to determine part location.
2. Procedure “C” must be done before removing the “old” board.
78
Keysight N5261/62A User’s & Service Guide
Содержание N5261A
Страница 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Страница 6: ......
Страница 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Страница 51: ...Keysight N5261 62A User s Service Guide 45 Installation System Operation Figure 1 30 Uncalibrated S12 and S21...
Страница 54: ...48 Keysight N5261 62A User s Service Guide...
Страница 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...
Страница 63: ...Service Information 2 Figure 2 8 Sloped Attenuator Cable Assembly Keysight N5261 62A User s Service Guide 57...