![Keysight Technologies N5261A Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/keysight-technologies/n5261a/n5261a_users-and-service-manual_1961871080.webp)
Service Information
3.
Set the PNA power output level to
get a power meter reading of 2 dBm ±
1dB.
4.
Connect the PNA Port 2
cable
to SRC1 RF IN
on the front panel
.
5.
Issue the Test Set I/O commands 32,0 and 0,1 (enable ALC and output to
Port 1).
6.
Measure the Port 1 RF OUT power using the power sensor. Confirm that RF
OUT is
+12.5 dBm (± 1.5 dB).
7.
Repeat the previous three steps for each of the remaining test set ports
using appropriate SRC1/2 connections and Test Set I/O commands.
Figure 2-
1
5
RF Output Gain with +2
dBm ALC On
Figure 2-
1
6
RF Output Gain with +2
dBm ALC Off
74
Keysight N5261/62A User’s & Service Guide
Содержание N5261A
Страница 1: ...User s and Service Guide Keysight Technologies N5261A and N5262A Millimeter Head Controller...
Страница 6: ......
Страница 7: ...Keysight N5261 62A User s Service Guide 1 Installation System Operation...
Страница 51: ...Keysight N5261 62A User s Service Guide 45 Installation System Operation Figure 1 30 Uncalibrated S12 and S21...
Страница 54: ...48 Keysight N5261 62A User s Service Guide...
Страница 55: ...Keysight N5261 62A User s Service Guide 49 Service Information...
Страница 63: ...Service Information 2 Figure 2 8 Sloped Attenuator Cable Assembly Keysight N5261 62A User s Service Guide 57...