36
Keysight N5261/62A User’s & Service Guide
Installation & System Operation
1-
Millimeter-wave Module Cable Connections
Before connecting the millimeter-wave modules, verify that the test set and the
power supplies (if used) are powered down.
There are four RF cables for each T/R module. Cables provided with the VDI
modules are pre-labeled. The cables for use with the OML modules come in
two types, two lighter and two heavier. Use the lighter cables for the IF signals
and the heavier cables for the RF and LO signals.
Figure 1-23
Example
-
OML Module Connections
Figure 1-24
Example
-
VDI Module Connections
DC Bias
REF IF
TEST IF
RF LO
REF IF
M EAS IF
R F
LO
I
O
Pow er Supply
Each module has
a separate power
supply.