Performance Verification
6
N2750/1/2A User’s Guide
61
Broadband Load
Keysight SMA male broadband load (Keysight P/N: 1250-3745). This load is
included
in the N5443A PV and deskew PV and deskew fixture for InfiniiMax
III. You can also purchase this load from Mini-circuits (manufacturer part
number is
ANNE-50X+
). An alternate broadband load is included in the
85056KE01 2.92 mm Standard Mechanical Calibration Kit, DC to 40 GHz.
?
3.5 mm adapter
Keysight 5062-1247 adapter, outside thread 3.5 mm (male) to 3.5 mm
(female).
?
Termination
BNC 50 ohm male terminator or equivalent (
not a critical part
). For example,
Pomona number 3840-50 or 4119-50.
?
Probe Positioner
Optional
. N2787A 3D Probe Positioner or equivalent.
?
Small Bench Vise
Optional
. For holding PV fixture.
?
Table 6
Required Test Equipment (Sheet 2 of 2)
Test Equipment
Recommended Model
Performance Test
1
2
Содержание N2750A
Страница 1: ...Keysight N2750 1 2A Differential Probes User s Guide ...
Страница 28: ...28 N2750 1 2A User s Guide 1 Using the Probes ...
Страница 34: ...34 N2750 1 2A User s Guide 2 SPICE Models Figure 16 SPICE Model ...
Страница 38: ...38 N2750 1 2A User s Guide 2 SPICE Models ...
Страница 42: ...42 N2750 1 2A User s Guide 3 Probe Calibration ...
Страница 54: ...54 N2750 1 2A User s Guide 5 Performance Data Plots Figure 35 Frequency Response Common Mode ...
Страница 56: ...56 N2750 1 2A User s Guide 5 Performance Data Plots Figure 38 Frequency Response Common Mode ...
Страница 66: ...66 N2750 1 2A User s Guide 6 Performance Verification Figure 45 Probe Connected to Probe Adapter not to scale ...
Страница 74: ...74 N2750 1 2A User s Guide 6 Performance Verification ...