Performance Verification
Figure 47
Close-Up Showing Tip Spacing and Position
g
Position the probe’s A tip on the center conductor as close to the edge of the
conductor as possible. Position the probe’s B tip to the ground, as close to
the dielectric as possible.
18
On the VNA, press
[Trigger] > Single
to trigger a single sweep. The display should
NOTE
If it looks noticeably different, the probe tip wires may not be making contact
under the pinchers.